I、Announcements
l     The equipment has high voltage, keep all the laps or boards closed.
l     When the equipment starts, put the heads on the hooks
l     The operator needs to wear protection glasses.
l     The eyes should not watch the light source (from where the  strong light shoot )
l     The treatment should not point at other parts of patient aimlessly
 
II、Operation skill and technique
 
1. Preparation before Treatment.
A)Beautician should make treatment file for customer , make full acknowledge of customer before treatment , then according to the situation of customer ,adjusting the energy of pulse ; Beautician should tell the customer of the merits of Regina and obtain his/her best coordination.
B) The treatment tools should be cleaned and sterilized before treatment. E.g. shaving the hair before hair removal.
C) The operator should wear white or light color cloth, wearing protection glasses.
2、Parameter setting
Remarks: Rule of Parameter adjustment:。Adjust the pulse width according to speckle color; adjust the pulse delay according to skin color. Mainly adjust the energy.
SR RULE : 1.Darker skin 、longer delay 、lower energy
          2.Whiter skin、 shorter delay、 higher energy
          3.Rougher vein、longer pulse width .
          4.Smoother vein、 shorter pulse width .
          5.Thicker vein、 longer pulse width; thinner vein 、shorter pulse width;
 HR RULE: 1.Darker hair, longer pulse width; Lighter hair, shorter pulse width.
          2. Thicker hair, longer delay, lower energy
          3. Thinner hair, shorter delay, higher energy
          4. Denser hair, lower energy; rarer hair, higher energy
          5. Darker skin, longer delay. Whiter skin, shorter delay, higher energy. 
1) Install and operate the machine according to our manual and directions.
2) The energy/ parameter adjustment changes according to the rule from low to high.
3) Energy/Parameter adjustment changes according the rule above.
4) The basic parameters should consider of the manual .and in the practical condition (Treatment) ,parameters should be changed according to the skin type 、body condition and the pain feeling of the patient .
5) IF any condition urgent in the treatment or about the technology, please contact us quickly .In case of hurting people and machine damage.
III、Treatment
 
1、Enter the working interface :
Connect the power line, open the key switch, and wait for about 3 seconds, Show as the Interface 2
2、Setting of treatment parameters:
We suggest to start from lower energy to make the treatment, firstly can make a facular test to watch the reaction of the patient.
3、Simmer the treatment head:
After the parameter setting in Interface 2 and 3, back to interface 2, and then press Simmer Key “STA”. We will hear tinkle of relay, the pulse light of treatment head will be bright at the moment.
(Notice , the interval of Pressing STA from last time should be over 4 seconds).
4、Smear the couplant on the test area, the IPL light will enter skin via couplant (cooling gel ) , so operator should do the treatment as long as the couplant smeared .
5、Vertically aim the treatment head at the skin , press slightly , make the treatment head full touch with couplant
6、Make the facular test , then clean the couplant , watching the reaction of the patient.
7、Reset the treatment parameter , start the treatment in the area smeared with couplant .
8、Press the button on the head , then will flash a intense pulse light , then the system will charge electricity automatically , the charge time will just need one second.
9、Clean the couplant and watch the reaction of the patient.
10、Treat next area , one mm. Overlap of last treatment area is permitted .
11、Make the record of the parameters used for the patient , which can be used as reference in next treatment .
Notice: After treatment, the patient should come back several weeks, to see whether need further treatment.
IV、Close-down
 
1、Under Interface 2 , firstly press “STA” ,then the light will be closed , then press “RF ” return to interface 1 , turn the key left , the machine will be close-down .
2、In the urgent condition , press urgent stop will stop the machine immediately . Turn the key clockwise, and then the machine can work again. If not, the machine can not work.
 